Team:University College London/Protocols/Nanodrop

From 2012.igem.org

(Difference between revisions)
(Nanodrop)
Line 1: Line 1:
{{:Team:University_College_London/templates/head|coverpicture=week4}}
{{:Team:University_College_London/templates/head|coverpicture=week4}}
-
<div class="protocol-nanodrop">== Nanodrop ==</div>
+
== Nanodrop ==
Software ND-1000
Software ND-1000

Revision as of 15:36, 16 July 2012

Nanodrop

Software ND-1000 Model:

1.Initialise the spectrophotometer by pipetting 1 µ of clean water onto lower optic surface, lowering the lever arm and selecting ‘initialise’ in the ND-1000 software

2.Wipe and add elution buffer as negative control. Click blank in ND-1000 software

3.Wipe and add 1 µl sample

4.On the software set lambda to 260nm

5.Lower the lever arm and click measure in ND-1000 software

6.Take readings for concentration and purity

7.Once measurement complete, wipe surface