Team:Wageningen UR/Protocol/Mutagenesis
From 2012.igem.org
Contents |
FPLC user manual
Every buffer that you put into the FPLC must be filtered and be void of air to optimize the separation.
Washing column
- Empty waste bin
- Set loop valve on Load Position
- Set column valve on designated position
- Set on the pressure alarm: 5mPa
- Set flow rate on half the maximum the column can handle
- Run the FPLC for about 2 times the column volume
- Empty waste bin
Washing loop
- Set flow rate on 0
- Set loop valve on Inject Position
- Set column valve on Bypass position
- Set on the pressure alarm: 5mPa
- Set flow rate on half the maximum the column can handle
- Run the FPLC for about 2 times the loop volume
Loading sample
- Set flow rate on 0
- Set loop valve on Load Position
- Inject the sample into the loop with an syringe
- Keep the syringe in the machine
Injecting sample onto the column
- Set flow rate on 0
- Set loop valve on Inject Position
- Set column valve on designated position
- Set on the pressure alarm: 5mPa
- Set flow rate on half the maximum the column can handle
- Run the FPLC for about 1 times the loop volume
- Turn on the Frac 900, to collect the samples
Washing column
- Set flow rate on 0
- Empty waste bin
- Set loop valve on Load Position
- Set column valve on designated position
- Set on the pressure alarm: 5mPa
- Set flow rate on half the maximum the column can handle
- Run the FPLC for about 2 times the column volume
- Empty waste bin