Team:Wageningen UR/Protocol/FPLC

From 2012.igem.org

Revision as of 07:28, 11 September 2012 by MarkvanVeen (Talk | contribs)

Contents

FPLC user manual

washing column

  1. Empty waste bin
  2. Set loop valve on Load Position
  3. Set column valve on designated position
  4. Set on the pressure alarm: 5mPa
  5. Set flow rate on half the maximum the column can handle
  6. Run the FPLC for about 2 times the column volume

washing loop

  1. Set flow rate on 0
  2. Set loop valve on Inject Position
  3. Set column valve on Bypass position
  4. Set on the pressure alarm: 5mPa
  5. Set flow rate on half the maximum the column can handle
  6. Run the FPLC for about 2 times the loop volume

loading sample

  1. Set flow rate on 0
  2. Set loop valve on Load Position
  3. Inject the sample into the loop with an syringe
  4. Keep the syringe in the machine

injecting sample onto the column

  1. Set flow rate on 0
  2. Set loop valve on Inject Position
  3. Set column valve on designated position
  4. Set on the pressure alarm: 5mPa
  5. Set flow rate on half the maximum the column can handle
  6. Run the FPLC for about 1 times the loop volume

washing column