FPLC user manual
washing column
- Empty waste bin
- Set loop valve on Load Position
- Set column valve on designated position
- Set on the pressure alarm: 5mPa
- Set flow rate on half the maximum the column can handle
- Run the FPLC for about 2 times the column volume
washing loop
- Set flow rate on 0
- Set loop valve on Inject Position
- Set column valve on Bypass position
- Set on the pressure alarm: 5mPa
- Set flow rate on half the maximum the column can handle
- Run the FPLC for about 2 times the loop volume
loading sample
- Set flow rate on 0
- Set loop valve on Load Position
- Inject the sample into the loop with an syringe
- Keep the syringe in the machine
injecting sample onto the column
- Set flow rate on 0
- Set loop valve on Inject Position
- Set column valve on designated position
- Set on the pressure alarm: 5mPa
- Set flow rate on half the maximum the column can handle
- Run the FPLC for about 1 times the loop volume