+ | = FPLC user manual = | ||
+ | '''Every buffer that you put into the FPLC must be filtered and be void of air to optimize the separation.''' | ||
+ | |||
+ | == Washing column == | ||
+ | <ol> | ||
+ | <li>Empty waste bin</li> | ||
+ | <li>Set loop valve on Load Position</li> | ||
+ | <li>Set column valve on designated position</li> | ||
+ | <li>Set on the pressure alarm: 5mPa</li> | ||
+ | <li>Set flow rate on half the maximum the column can handle</li> | ||
+ | <li>Run the FPLC for about 2 times the column volume</li> | ||
+ | <li>Empty waste bin</li> | ||
+ | </ol> | ||
+ | |||
+ | == Washing loop == | ||
+ | <ol> | ||
+ | <li>Set flow rate on 0</li> | ||
+ | <li>Set loop valve on Inject Position</li> | ||
+ | <li>Set column valve on Bypass position</li> | ||
+ | <li>Set on the pressure alarm: 5mPa</li> | ||
+ | <li>Set flow rate on half the maximum the column can handle</li> | ||
+ | <li>Run the FPLC for about 2 times the loop volume</li> | ||
+ | </ol> | ||
+ | |||
+ | == Loading sample == | ||
+ | <ol> | ||
+ | <li>Set flow rate on 0</li> | ||
+ | <li>Set loop valve on Load Position | ||
+ | <li>Inject the sample into the loop with an syringe | ||
+ | <li>Keep the syringe in the machine | ||
+ | </ol> | ||
+ | |||
+ | == Injecting sample onto the column == | ||
+ | <ol> | ||
+ | <li>Set flow rate on 0</li> | ||
+ | <li>Set loop valve on Inject Position</li> | ||
+ | <li>Set column valve on designated position</li> | ||
+ | <li>Set on the pressure alarm: 5mPa</li> | ||
+ | <li>Set flow rate on half the maximum the column can handle</li> | ||
+ | <li>Run the FPLC for about 1 times the loop volume</li> | ||
+ | <li>Turn on the Frac 900, to collect the samples</li> | ||
+ | </ol> | ||
+ | |||
+ | == Washing column == | ||
+ | <ol> | ||
+ | <li>Set flow rate on 0</li> | ||
+ | <li>Empty waste bin</li> | ||
+ | <li>Set loop valve on Load Position</li> | ||
+ | <li>Set column valve on designated position</li> | ||
+ | <li>Set on the pressure alarm: 5mPa</li> | ||
+ | <li>Set flow rate on half the maximum the column can handle</li> | ||
+ | <li>Run the FPLC for about 2 times the column volume</li> | ||
+ | <li>Empty waste bin</li> | ||
+ | </ol> |
Latest revision as of 07:40, 11 September 2012
Contents |
FPLC user manual
Every buffer that you put into the FPLC must be filtered and be void of air to optimize the separation.
Washing column
- Empty waste bin
- Set loop valve on Load Position
- Set column valve on designated position
- Set on the pressure alarm: 5mPa
- Set flow rate on half the maximum the column can handle
- Run the FPLC for about 2 times the column volume
- Empty waste bin
Washing loop
- Set flow rate on 0
- Set loop valve on Inject Position
- Set column valve on Bypass position
- Set on the pressure alarm: 5mPa
- Set flow rate on half the maximum the column can handle
- Run the FPLC for about 2 times the loop volume
Loading sample
- Set flow rate on 0
- Set loop valve on Load Position
- Inject the sample into the loop with an syringe
- Keep the syringe in the machine
Injecting sample onto the column
- Set flow rate on 0
- Set loop valve on Inject Position
- Set column valve on designated position
- Set on the pressure alarm: 5mPa
- Set flow rate on half the maximum the column can handle
- Run the FPLC for about 1 times the loop volume
- Turn on the Frac 900, to collect the samples
Washing column
- Set flow rate on 0
- Empty waste bin
- Set loop valve on Load Position
- Set column valve on designated position
- Set on the pressure alarm: 5mPa
- Set flow rate on half the maximum the column can handle
- Run the FPLC for about 2 times the column volume
- Empty waste bin