Team:TU Darmstadt/Protocols/GC-MS/GC-Protocol

From 2012.igem.org

Revision as of 19:26, 25 September 2012 by S jager (Talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)

Contents

INSTRUMENT CONTROL PARAMETERS

Control Information

  • Sample Inlet  : GC
  • Injection Source  : GC ALS
  • Mass Spectrometer : Enabled

Oven

  • Equilibration Time 0.25 min
  • Oven Program On
   55 °C for 0.5 min
   then 30 °C/min to 250 °C for 3 min
  • Run Time 10 min

Front Injector

  • Syringe Size 10 µL
  • Injection Volume 1 µL
  • Injection Repetitions 1
  • Solvent A Washes (PreInj) 2
  • Solvent A Washes (PostInj) 2
  • Solvent A Volume 5 µL
  • Solvent B Washes (PreInj) 2
  • Solvent B Washes (PostInj) 2
  • Solvent B Volume 5 µL
  • Sample Washes 1
  • Sample Wash Volume 1 µL
  • Sample Pumps 2
  • Dwell Time (PreInj) 0 min
  • Dwell Time (PostInj) 0 min
  • Solvent Wash Draw Speed 300 µL/min
  • Solvent Wash Dispense Speed 6000 µL/min
  • Sample Wash Draw Speed 300 µL/min
  • Sample Wash Dispense Speed 6000 µL/min
  • Injection Dispense Speed 6000 µL/min
  • Viscosity Delay 1 sec
  • Sample Depth Disabled

Back Injector

  • Front SS Inlet He
  • Mode Pulsed Splitless
  • Heater On 270 °C
  • Pressure On 0.54761 bar
  • Total Flow On 54 mL/min
  • Septum Purge Flow On 3 mL/min
  • Gas Saver On 15 mL/min After 3 min
  • Injection Pulse Pressure 1 bar Until 0.5 min
  • Purge Flow to Split Vent 50 mL/min at 2 min


Thermal Aux 1 {MSD Transfer Line}

  • Heater On
  • Temperature Program On
   320 °C for 0 min
  • Run Time 10 min


Column #1

  • HP-5MS 5% Phenyl Methyl Silox: 0001
  • HP-5MS 5% Phenyl Methyl Silox
  • 325 °C: 30 m x 250 µm x 0.25 µm
  • In: Front SS Inlet He
  • Out: Vacuum
  • (Initial) 55 °C
  • Pressure 0.54761 bar
  • Flow 1 mL/min
  • Average Velocity 36.534 cm/sec
  • Holdup Time 1.3686 min
  • Flow Program Off
  • 1 mL/min for 0 min
  • Run Time 10 min